Carrier 58GC Specifiche Pagina 126

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 140
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 125
7820 MSD Site Preparation
126 GC, GC/MS, and ALS Site Preparation Guide
filters are designed for use with the GC, GC/MS, ICP- OES, ICP- MS,
LC/MS, and any other analysis instrument using carrier gas. Six filters are
available, including CO2, oxygen, moisture, and organics trap (charcoal).
Filter types
Each Gas Clean Filter type is designed to filter out a specific impurity
that may exist in the gas supply. The following filter types are available:
•Oxygen - Prevents oxidation of the GC column, septum, liner, and glass
wool.
•Moisture - Delivers fast stabilization times for increased GC
productivity, and prevents hydrolization damage to the stationary phase,
column, liner, glass wool, or septum in the GC.
Process Moisture - Prevents oxidation of GC components and is safe to
use with acetylene in process GC applications.
Charcoal - Removes organic compounds and ensures correct
performance of FID detectors in the GC.
•GC/MS - Delivers fast stabilization times for increased GC productivity,
removes oxygen, moisture, and hydrocarbons from the carrier gas for
MS applications, and provides ultimate GC column protection.
Table 64 on page 127 shows recommended filter connection diagrams for
common instrument configurations.
Vedere la pagina 125
1 2 ... 121 122 123 124 125 126 127 128 129 130 131 ... 139 140

Commenti su questo manuale

Nessun commento