7890 Series GC Site Preparation
GC, GC/MS, and ALS Site Preparation Guide 23
USA fast heating oven, 240 V
The 240 V fast heating oven requires 240 V/15 A power. Do not use 208 V
power. Lower voltage causes slow oven ramps and prevents proper
temperature control. The power cord supplied with your GC is rated for
250 V/15 A, and is a two- pole, three- wire cord with grounding (type
L6- 15R/L6- 15P).
Canadian installation
When installing a GC in Canada, make sure your GCs power supply
circuit meets the following additional requirements:
The circuit breaker for the branch circuit, which is dedicated to the
instrument, is rated for continuous operation.
The service box branch circuit is marked as a “Dedicated Circuit.”
Common instrument power cord plugs
Table 4 below shows common Agilent power cord plugs.
Table 4 Power cord terminations
Country
Voltage
Amps
Cable
length (m)
Wall termination
Plug
Termination
Australia 240 16 2.5 AS 3112
China 220 15 4.5 GB 1002
Vedere la pagina 22