7890 Series GC Site Preparation
48 GC, GC/MS, and ALS Site Preparation Guide
filters are designed for use with the GC, GC/MS, ICP- OES, ICP- MS,
LC/MS, and any other analysis instrument using carrier gas. Six filters are
available, including CO2, oxygen, moisture, and organics trap (charcoal).
Filter types
Each Gas Clean Filter type is designed to filter out a specific impurity
that may exist in the gas supply. The following filter types are available:
•Oxygen - Prevents oxidation of the GC column, septum, liner, and glass
wool.
•Moisture - Delivers fast stabilization times for increased GC
productivity, and prevents hydrolization damage to the stationary phase,
column, liner, glass wool, or septum in the GC.
Process Moisture - Prevents oxidation of GC components and is safe to
use with acetylene in process GC applications.
Charcoal - Removes organic compounds and ensures correct
performance of FID detectors in the GC.
•GC/MS - Delivers fast stabilization times for increased GC productivity,
removes oxygen, moisture, and hydrocarbons from the carrier gas for
MS applications, and provides ultimate GC column protection.
Table 24 on page 49 shows recommended filter connection diagrams for
common instrument configurations.